KALIBRASI FLOW METER DALAM ALIRAN FLUIDA PADA SISTEM MANIFOLD

CALIBRATION OF FLOW METER IN FLOW RATE MANIFOLD SYSTEM

Author :
ASY’ARI, MUHAMMAD KHAMIM ( 2411031018 )




ABSTRAK

Telah dirancang alat eksperimen kalibrasi flow meter pada sistem manifold yang berfungsi untuk mengetahui nilai ketidakpastian serta untuk mengetahui pengaruh sistem manifold terhadap kalibrasi flow meter. Cara kerja sistem alat eksperimen ini yaitu pada kondisi awal ditentukan besar bukaan manual valve. Kemudian ditekan push button sehingga solenoid valve terbuka dan mengalirkan air ke talang air. Selanjutnya air mengalir melalui flow meter. Air keluaran flow meter akan ditampung dalam gelas ukur dihitung nilai volumenya dan digunakan sebagai pembacaan standar. Sedangkan perubahan nilai dari flow meter digunakan sebagai pembacaan alat. Dari hasil kalibrasi flow meter didapatkan nilai ketidakpastian diperluas terkecil untuk valve dengan bukaan pada flow meter B sebesar 00111 L untuk valve dengan bukaan pada flow meter B sebesar 00978 L untuk valve dengan bukaan pada flow meter D sebesar 00741 L dan untuk valve dengan bukaan 1 pada flow meter D sebesar 01172 L.


ABSTRACT

Device experiment calibration of flowmeter in flowrate manifold system has been designed which used to determine the values of the uncertainty and to determine the effect manifold system of calibration flowmeter. Work principle of this experimental device is in the first conditions has been determined the large of valve opening. Then the push button is pressed so the solenoid valve is open and drain the water in to water storage. Then the water drain through the flowmeter. Water flowmeter output will be accomodated in a glass measuremen. Whereas the change value of flowmeter uses as device reading. The result of this flowmeter calibration the values obtained that the minimum expanded uncertainty of valve in opening in flowmeter B is 00111 L the expanded uncertainty of valve in opening in flowmeter B is 00978 L the expanded uncertainty of valve in opening in flowmeter B is 00741 L the expanded uncertainty of valve in 1 opening in flowmeter B is 01172 L.



Keywordsflow meter, kalibrasi, ketidakpastian diperluas, sistem manifold
 
Subject:  Kalibrasi
Contributor
  1. Dr. Ir. Ali Musyafa’ , M.Sc.
Date Create: 22/09/2015
Type: Text
Format: PDF
Language: Indonesian
Identifier: ITS-NonDegree-24102150001257
Collection ID: 24102150001257
Call Number: RSF 620.106 4 Asy k


Source
Non Degree Of Physics Engineering RSF 620.106 4 Asy k, 2015

Coverage
ITS Community

Rights
Copyright @2015 by ITS Library. This publication is protected by copyright and per obtained from the ITS Library prior to any prohibited reproduction, storage in a re transmission in any form or by any means, electronic, mechanical, photocopying, reco For information regarding permission(s), write to ITS Library




[ Download - Open Access ]

  1.  ITS-NonDegree-38955-2411031018-abstract_id.pdf - 86 KB
  2.  ITS-NonDegree-38955-2411031018-abstract_en.pdf - 85 KB
  3.  ITS-NonDegree-38955-2411031018-conclusion.pdf - 94 KB




 Similar Document...




! ATTENTION !

To facilitate the activation process, please fill out the member application form correctly and completely

Registration activation of our members will process up to max 24 hours (confirm by email). Please wait patiently

POLLING

Bagaimana pendapat Anda tentang layanan repository kami ?

Bagus Sekali
Baik
Biasa
Jelek
Mengecewakan





You are connected from 54.80.26.116
using CCBot/2.0 (http://commoncrawl.org/faq/)



Copyright © ITS Library 2006 - 2017 - All rights reserved.
Dublin Core Metadata Initiative and OpenArchives Compatible
Developed by Hassan